Condition
Battery Low
Description and Corrective Action
The Battery Low warning and alarm will only be
activated when the pump is operating from the
battery (the mains power is disconnected).
When any of these Battery Low warning or alarm
screens are activated, connect the pump to the
mains power supply to recharge the battery and
continue therapy.
If the pump is in Standby:
• The warning will be activated when the remaining
charge in the battery is less than 15%. The
battery icon flashes. Therapy can be started.
Connect to the mains power supply to clear the
warning.
• When the charge in the battery is less than 10%,
therapy cannot be started.
Connect to the mains power supply to clear the
warning and start therapy.
If the pump is in the Run state:
• The alarm will be activated when the remaining
charge in the battery is less than 10%. There is
an audible tone and the battery icon flashes. The
LED indicators on the front and underside of the
pump change to yellow.
Therapy can continue. Pressing the Left Control
button will silence the audible tone.
Connect to the mains power supply to clear the
warning and continue therapy.
• If the audible alarm is muted when the remaining
charge in the battery is less than or equal to 7%,
then an audible tone will sound and the LED
indicator will change to yellow. The audible tone
can no longer be silenced. The pump should be
connected to the mains power supply
immediately.
• When the remaining charge in the battery is
less than 5%, the pump will shut down with
no additional visible or audible indications.
24