Low Pressure/Leak
Kinked/Blocked Tube
Garment Disconnected
Service Required
Description and Corrective Action
The warning is activated after 4 minutes and shows
a leak in garment 1 or its tubing.
The warning changes to an alarm after an additional
6 minutes (10 minutes total).
Examine the garment and tubing for leaks. The
warning or alarm will be cleared if the leak is
repaired. If the alarm continues, replace the affected
garment.
The warning is activated after 4 minutes and shows
garment 1 has a kinked or blocked tube.
The warning changes to an alarm after an additional
6 minutes (10 minutes total).
Examine the garment and tubing for kinks or
blockages. The warning or alarm will be cleared if
the kink or blockage is repaired. If the alarm
continues, replace the affected garment.
This warning is activated after 1 minute if the pump
detects that a garment has been disconnected while
the pump is in the Run state; it shows that garment 1
has been disconnected. The "garment type" icon
flashes.
The warning changes to an alarm after an additional
9 minutes (10 minutes total) if the pump continues to
detect that a garment has been disconnected while
the pump is in the Run state; it shows that garment 1
has been disconnected.
Reconnect or replace the garment. The warning or
alarm will be cleared if the garment is reconnected.
The service "spanner" icon appears at the right side
of the screen to indicate that the pump requires
service attention. This will normally occur after the
preset service interval has expired.
The LED indicators on the front and underside of the
pump remain green.
Therapy can continue until service is available.
Call the service engineer.