Garment Type and
Inflation Indication
Uniform (DVT) Calf
(or Calf & Thigh)
Sequential (Tri Pulse) Calf
(or Calf & Thigh)
Tubeset Connector
Number
Start/Stop Therapy
Garment Pressure
Indication
These show which garment type is connected to
each tubeset and when each garment is being
inflated.
A garment connected but not inflated is shown as
an outline on the leg icon. When it is inflated, the
garment outline is filled in black.
Garment
Foot
The garment connector on the end of each tubeset
has a push button which is colour-coded and has a
number marked on it: blue "1" or orange "2". The
numbers correspond with the "1" and "2" on the left
side of the LCD screen adjacent to the leg icons.
The icon above the Middle Control button changes
between Run and Standby modes, as follows.
Start Therapy icon. In normal operation, this
icon is displayed when the pump is in
Standby and at least one garment is
connected. Press the button to start therapy.
Stop Therapy icon. In normal operation, this
icon is displayed when the pump is in Run
mode. Press the button to stop therapy.
The default target inflation pressure for each garment
is as follows:
• Foot: 130 mmHg.
• Uniform calf and calf & thigh: 40 mmHg.
• Sequential calf and calf & thigh: 45 mmHg.
8