SYMBOL DEFINITIONS
Danger - No smoking near patient or
device.
This symbol has a red circle and
diagonal bar on the product label.
It is mandatory to read and understand the
operating instructions prior to use.
This symbol has a blue background on
the product label.
Use no Oil, Grease or Lubricants
This symbol has a red circle and
diagonal bar on the product label.
RTCA / DO-160G Section 21 Category M
and Section 20 Category T - The
manufacturer of this POC has determined
this device conforms to all applicable FAA
acceptance criteria for POC carriage and
use on board aircraft.
Text shown with this symbol is red on
the product label.
Do not use near heat or open flames.
This symbol has a red circle and
diagonal bar on the product label.
Do Not Disassemble
This symbol has a red circle and
diagonal bar on the product label.
Direct Current (DC Power)
Class II electrical protection – double
insulated
Type BF applied part
CAUTION: Federal (U.S.A.) law restricts this device to sale by or on the order of a
physician.
IP22 Ingress Protection - Protected against finger access to hazardous parts; protected
IP22
against vertically falling water drops when enclosure is tilted up to 15°
This device contains electrical and/or electronic equipment that must be recycled per
EU Directive 2012/19/EU- Waste Electrical and Electronic Equipment (WEEE)
BATTERY PACK MARKINGS
Attention, Consult Instruction Guide
Refer to Operating Instructions
Mark of conformity to applicable European
Directives
Regulatory Compliance Mark of Australia and
New Zealand.
Battery Charge Status Gauge
IMPORTANT SAFEGUARDS
Read this entire guide before using your Drive DeVilbiss iGo2 portable oxygen concentrator. Important safeguards are indicated throughout this guide; pay special attention to all safety
information. Imminently and potentially hazardous information is highlighted by these terms:
DANGER
Indicates an imminently hazardous situation which could result in death or serious injury to the user or operator if not avoided.
WARNING
Indicates a potentially hazardous situation which could result in death or serious injury to the user or operator if not avoided.
EN - 4
Catalog Number
Serial Number
No MR (MR Unsafe) Unsafe
for Magnetic Resonance
Environment
TUV Rheinland
C
US
Manufacturing Lot Code
Manufacturer
European Representative
European CE Mark
Operating Temperature
+35˚C
Range +15 to +35˚C (+41 to
+95˚F)
+5˚C
Storage & Transport
+60˚C
(140˚F)
Temperature Range -25˚C to
-25˚C
+60˚C (-13˚F to +140˚F)
(-13˚F)
UL Recognized for Canadian and US
market
Recycling Symbol Taiwan
UN Transportation Test
Recycling Symbol
Li-ion
Power Button
Increase Prescription
Setting Button
Decrease Prescription
Setting Button
Bluetooth connected to a
smartphone (125X-BT
Series Bluetooth Models
Only)
Bluetooth not connected to
a smartphone (125X-BT
Series Bluetooth Models
Only)
SmartDose Enabled
Alarm Silenced button -
Symbol Illuminates on
display when an active
alarm is silenced
Hour Meter
93%
Storage & Transport
Humidity Range 15-93%
15%
Medical Device
General Warning
This symbol is used throughout this manual to indicate hazardous
situations to avoid.
Important Information
This symbol is used throughout this manual to indicate important
information you should know.
Note and Information Symbol
This symbol is used throughout this manual to indicate notes, useful
tips, recommendations and information.
Dispose of this product according to local regulations.
For Canada & USA: Please call 1-800-822-8837 for
information on how to recycle this battery
China RoHS
Regulatory Compliance Mark of Japan
Green Oxygen Output within
Specification
Yellow Low Oxygen, less than
86% O
2
Yellow Startup Low Oxygen,
less than 86% O
2
Red Service Required Symbol
Red High Temperature
(> 55 ˚C)
Battery Charge Remaining
This battery symbol is illuminated with
the following colors:
Red Bar Graph when
5%
≤ 5% battery charge
Yellow Bar Graph when
10%
≤ 10% battery charge
50%
Green Bar Graph when
80%
10% - 100% battery charge
100%
Battery Charging
Lightning bolt in the
50%
battery tip indicates battery
is charging.
External Power Connected
A-125B