LC300 Menu Functions
Use the rotary switch and the push-buttons to set the display and programming values for your
application.
Rotary
Switch 
0
Position
Read PV
Display
(pF)
Up button
Down 
button
Both 
buttons
Maintenance
SITRANS LC300 requires no maintenance or cleaning under normal operating conditions. Even
with significant build-up on the SITRANS LC300 probe, the level controller will continue to
operate.
Build-up of material on the active shield area has little or no effect on the
Note:
performance of the SITRANS LC300.
Unit Repair and Excluded Liability
For detailed information, please see the inside back cover.
Page EN-10
SITRANS LC300 – QUICK START MANUAL
All manuals and user guides at all-guides.com
1
2
Read LRV (pF)
Read URV (pF)
(0% level)
(100% level)
Increase LRV
Increase URV
Decrease LRV
Decrease URV
Set LRV from PV Set URV from PV
3
4
Read mA loop-
Diagnostics
current
Product 
Set fault protection
setting to 22 mA
Version
Set fault protection
setting to 3.6 mA
Reset/
Disable fault
Acknowledge
protection
Fault
5
Damping
Increase
Damping
Decrease
Damping
Set Damping
to 1.00