Chuck cleaning
Clean Push Button chuck once a week.
1) Mount the Tip Nozzle into the spray can port.
2) Lubricate the chuck directly through the bur insertion hole. (Fig.14)
3) Lubricate the handpiece by using NSK PANA SPRAY Plus (Fig.11,13)
or NSK automatic handpiece cleaning and lubrication system.
CAUTION
If the chuck is not regularly cleaned the chuck grip may be
weakened and the bur may be accidentally released while in
operation.
NSK automatic handpiece cleaning and lubrication system
When using NSK automatic handpiece cleaning and lubrication
system, refer to the system instructions.
7-4 Sterilization
Sterilize the product by autoclave sterilization. Remove the bur after
each patient and sterilize as noted below.
1) Insert into an autoclave pouch. Seal the pouch.
2) Autoclavable under the conditions below.
Autoclave for more than 20 min. at 121°C, or 15 min. at 132°C, or
3 min. at 134°C.
3) The handpiece should remain in the autoclave pouch until required
for use.
CAUTION
Do not autoclave the product with other instruments even when it
is in a pouch. This is to prevent possible discoloration and damage
to the product from chemical residue on other instruments.
Keep the product in suitable atmospheric pressure, temperature,
humidity, ventilation, and sunlight. The air should be free from
dust, salt and sulphur.