  • FRANÇAIS, page 30
7-3 Lubrication
NSK PANA SPRAY Plus
Pana-Max2 M4/Pana-Max2 B2
Apply NSK PANA SPRAY Plus every time after each use and/or before
autoclaving.
1) Remove the handpiece from the hose.
2) Remove the bur from the handpiece.
3) Mount the Tip Nozzle into the spray can port. (Fig.11)
4) Insert the Tip Nozzle into the Drive Air Port of the handpiece. Hold
the handpiece and spray for approximately 2-3 seconds. Apply
lubricant until it expels from the handpiece head for at least 2
seconds. (Fig.12)
Pana-Max2 QD/Pana-Max2 PTL/Pana-Max2 KV
Apply NSK PANA SPRAY Plus every time after each use and/or before
autoclaving.
1) Remove the bur from the handpiece.
2) Insert the Spray Nozzle into the Spray Port nozzle on the can.
3) Insert the Spray Nozzle in rear of the handpiece. Hold the
handpiece and spray for approximately 2-3 seconds. Apply
6
lubricant until it expels from the handpiece head for at least 2
seconds. (Fig.13)
CAUTION
When applying spray be sure to hold the handpiece firmly to
prevent the handpiece from slipping out of the hand due to the
spray pressure.
Hold the spray can upright.