3.3.4
TEMPERATURE PROBE CALIBRATION
Connect the probe to the instrument as indicated in the electrical connections.
Select the TEMP. probe from the Calibration menu.
TEMP
CAL.
25.0°C
CL
CAL.
28.0°C
CL
CAL.
28.0°C
Wait
10"
Type: MAN
Type: MAN
Type: MAN
0000137242
Rev.3.1
In Automatic (MAN) mode:
Use a reference instrument to read the
temperature value.
Adjust the value shown on the display to
match the value read by the reference
instrument. Press Enter to confirm.
Wait 10 seconds for the calibration to
complete.
Once the operation has concluded, a
message will appear indicating that the
calibration has been carried out successfully.