3.3.4
TEMPERATURE PROBE CALIBRATION
Connect the probe to the instrument as indicated in the electrical connections.
Select the TEMP. probe from the Calibration menu.
TEMP
CAL.
25.0°C
CL
CAL.
28.0°C
CL
CAL.
28.0°C
Wait
10"
Type: MAN
Type: MAN
Type: MAN
0000137242
Rev.3.1
In Automatic (MAN) mode:
•
Use a reference instrument to read the
temperature value.
•
Adjust the value shown on the display to
match the value read by the reference
instrument. Press Enter to confirm.
•
Wait 10 seconds for the calibration to
complete.
•
Once the operation has concluded, a
message will appear indicating that the
calibration has been carried out successfully.
12