3. Then use this softkey to choose the desired edge direction: Trig Out
SCPI Command
OUTPut:TRIGger:SLOPe {POSitive|NEGative}
OUTPut:TRIGger {ON|1|OFF|0}
System-Related Operations
This section covers instrument state storage, power-down recall, error conditions, self test, and display control.
Though unrelated to waveform generation, these operations are important for instrument operation.
Instrument State Storage
– There are two ways to store and retrieve instrument states:
– Named state files, using the front panel or MMEMory:STORe:STATe and MMEMory:LOAD:STATe
– Memory locations 0 through 4, using *SAV and *RCL
– Both state storage methods remember the selected function (including arbitrary waveforms), frequency,
amplitude, DC offset, duty cycle, symmetry, and modulation parameters.
– Stored states are not affected by *RST; a stored state remains until overwritten or specifically deleted.
Front Panel Operations
See
Store or Retrieve the Instrument
Keysight EDU33210 Series User's Guide
State.
Off
| (Up) | (Down).
145