Inficon GAS-Mate Mode D'emploi page 9

Masquer les pouces Voir aussi pour GAS-Mate:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

Troubleshooting
W A R N I N G
Substitution of components may impair intrinsic safety.
Except for the batteries and the sensor, the internal parts of the GAS-Mate
Combustible Gas Leak Detector are not user serviceable. If you experience a
problem with your GAS-Mate, follow the steps in the trouble shooting guide below
to determine how to remedy your problem. If you cannot remedy the problem, take
your GAS-Mate to your wholesaler for warranty evaluation or call our technical
assistance line at +49 221 56788-660.
Problem
GAS-Mate will not power
up.
Poor sensitivity. Unit
powers up, but has a
difficult time detecting
combustible gas.
Alarm sounds
continuously, even after
a couple minutes of
warm-up.
Possible Cause
Batteries are worn out.
Batteries are improperly
installed.
Unit set to low sensitivity.
Sensor is worn out and
needs to be replaced.
The sensor is not
installed properly or is
missing.
Sensor is worn out and
needs to be replaced.
9
Remedy
Replace batteries.
Refer to
Figure 1
proper installation.
Increase sensitivity by
dragging finger upwards
on touch pad.
Replace sensor.
The sensor leads must
be straight and inserted
in the holes at the base
of the sensor socket. Or,
the orientation tab is not
properly aligned in the
probe tip. Refer to
2..
Replace sensor.
for
Figure

Publicité