LPS Series User Instructions
Place the test weight on the scale (between 25%-100% of full capacity for 2
point calibration; between 12.5% - 100% for three point calibration). Enter the
test weight value by using the HOLD and UNITS keys. Press the HOLD key to
select which digit to change (it will flash). Press the UNITS key to change the
flashing digit. Press the TARE key to accept this value.
The display will blink and then CAL.P2 is temporarily displayed, if
successful. Then xxx is displayed (100% of capacity is the default value)
with the last digit blinking. This shows the scale is ready for the P2 (25%-
100% of full capacity) calibration point.
Go to step
11
to finish the three point calibration, or, if you want to do a
two point calibration, leave the test weight on the scale, enter the same
test weight value that you entered above and press the TARE key.
CAL.P0 is temporarily displayed if successful, then 0.0 is displayed.
Skip to step 12.
CAL.eR is displayed if calibration failed. Repeat from step 8b.
Place the P2 weight on the scale and use the HOLD and UNITS keys to enter
this weight. Press the TARE key to accept. . .
The display will blink and then CAL.P0 is temporarily displayed if
successful, then 0.0 is displayed.
If calibration failed, repeat from step 8b.
Remove the weight from the scale and press the TARE key. . .
The display blinks and the zero point is captured again. If calibration is
successful the unit will go through the power up display routine. If an
error occurs in calibration, CAL.Er is temporarily displayed and the
indicator repeats the procedure from step 8b.
The indicator is now calibrated and in normal weighing mode.
19