3.4.1 Key Details
HOLD
1.
Note: Only stable weight can be held! The weight is stable when the unit of measure
symbol is steady. When
scale. When
If you press TARE when the hold function is enabled, the held reading will be cleared,
but the hold function is still enabled. To disable hold function, press HOLD once more.
2.
UNIT
1.
2.
TARE
1.
12
From normal weighing mode, press this key for less than five seconds to enable
or disable the Display Data Hold function. When this is enabled the
or continuously appear in the upper right of the display window.
There are three Hold Function modes (set these using instructions found
in section 3.5):
If Hld.00 is set, a stable displayed weight, above 10 divisions, will be held/
l
locked when the weight is removed from the scale. If you place a heavier
weight on the scale, that new weight will be displayed and held. If you
remove the weight and let the scale become stable, the held weight is still
displayed until you place a smaller weight on the scale. That smaller weight
will be displayed and held.
If Hld.01 is set, a stable displayed weight, above 10 divisions, will be held/
l
locked when the weight is removed from the scale. If you place a heavier
weight on the scale, that new weight will be displayed and held. Only a larger
weight than the held weight will change the display. This is different than the
Hld.00 mode.
If Hld.02 (... Hld.99) is set, the stable, displayed weight will be held/locked
l
as long as weight changes stay within the division window (± 2 . . . ±99) you
choose.
flashes, the displayed reading is the current weight on the
is continuously lit, the displayed reading is the held/locked reading.
Press this key for more than five seconds to access the Display Internal Code
mode. See section
3.7
Display Internal Code for more information.
Press this key for less than five seconds to toggle the unit of measure between
lb and kg. This information is saved when you turn off the scale. It will power up
in the same unit of measure.
Press this key for more than five seconds to access the Setup mode. In this
mode you can set the auto-off timer and select the serial communication
protocol.
From normal weighing mode, press this key less than five seconds to zero or
tare the scale if the weight is stable. When the current gross weight is in the
zero-point range, this will zero the scale and clear the NET. When current gross
weight is not in zero-point range, this will tare the scale and light NET on the
display window. If the weight is not stable, the scale will attempt the zero/tare
function for approximately 10 seconds.
will flash
LPS Series User Instructions