Guidance and Manufacturer's Declaration - Electromagnetic Immunity
Immunity Test
Note: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption, and reflection from structures, objects and people.
Warranty
Medtronic MiniMed, Inc. (or such other legal entity as may be referred to as manufacturer on the
labeling of this device "Medtronic MiniMed") warrants the Medtronic transmitter to the purchaser of
the product against defects in material and workmanship for a period of one (1) year and the charger
for up to one (1) year from the date of purchase.
During the warranty period, Medtronic MiniMed will replace or repair, at its discretion, any defective
transmitter or charger, subject to the conditions and exclusions stated herein. This warranty applies
only to new devices. In the event a transmitter or charger is replaced, the warranty period will not be
extended past its original expiration date.
This warranty is valid only if the Medtronic transmitter or charger is used in accordance with the
manufacturer's instructions. Without limitation, this warranty will not apply:
IEC 60601-1-2:2014 Test
Level
80% AM at 1 kHz
Max foreseeable
use condition
per IEC
60601-1-2:2014
80% AM at 1 kHz
40
Electromagnetic Environ‐
ment Guidance
should be used no closer to
any part of the transmitter
than the recommended sep‐
aration distance of 30 cm
(12 in).
Field strengths from fixed RF
transmitters, as determined
by an electromagnetic site
survey, should be less than
the compliance level in each
frequency range.
Interference may occur in
the vicinity of equipment
marked with the following
symbol: