11. Prepare a 1:10 bleach solution by using one (1) part 8.25% bleach to nine (9) parts water, for a final
concentration of 0.8%. Make sure to prepare a fresh solution for each use.
12. Make sure that the previous cleaning steps are completed before disinfection. With the tester
still attached, soak the transmitter in the bleach solution for 20 minutes.
13. Rinse the transmitter under running room temperature tap water for three minutes.
14. Place the transmitter and tester on a clean, dry cloth and air dry them completely.
WARNING: If there is any body fluid inside the connector opening on earlier inspection,
the transmitter must be discarded with the tester still attached, according to local
regulations for battery disposal (non-incineration).
15. Disconnect the tester from the transmitter by gently squeezing the arms of the tester.