Guidance and Manufacturer's Declaration - Electromagnetic Immunity
Immunity Test
IEC 60601-1-2:2014 Test
Electrostatic dis‐
±8 kV contact
charge (ESD)
±2 kV, ±4 kV, ±8 kV,
IEC 61000-4-2
±15 kV air
Conducted distur‐
3 V
bances induced
150 kHz to 80 MHz
by RF fields
6 V
ISM bands between
150 kHz to 80 MHz
Electrical fast tran‐
±2 kV 100 kHz repetition fre‐
sient/burst
quency
IEC 61000-4-4
Surge
Line to Line: ±0.5 kV, ±1 kV
IEC 61000-4-5
Line to Ground: ±0.5 kV,
±1 kV, ±2 kV
Note: U
is the a.c. mains voltage prior to application of the test level.
T
Voltage dips, short
0% U
interruptions, and
90°, 135°, 180°, 225°, 270°,
voltage variations
and 315°)
on power supply
0% U
lines
70% for 25/30 cycles (at 0°)
IEC 61000-4-11
0% for 250/300 cycles
Power frequency
30 A/m
(50/60 Hz) mag‐
netic field
IEC 61000-4-8
Proximity fields
IEC 60601-1-2:2014,
from RF wireless
Table 9
communications
equipment
IEC 61000-4-3
Note: U
is the a.c. mains voltage prior to application of the test level.
T
Radiated RF
10 V/m
IEC 61000-4-3
80 MHz to 2.7 GHz
Level
RMS
RMS
; 0.5 cycles (at 0°, 45°,
T
; 1 cycle (at 0°)
T
Max foreseeable
use condition
per IEC
60601-1-2:2014
±8 kV contact
For use in a typical domestic,
±2 kV, ±4 kV, ±8 kV,
commercial, or hospital envi‐
±15 kV air
ronment.
Not applicable
Requirement does not apply
to this battery powered
device.
Not applicable
Requirement does not apply
to this battery powered
device.
Not applicable
Requirement does not apply
to this battery powered
device.
Not applicable
Requirement does not apply
to this battery powered
device.
30 A/m
For use in a typical domestic,
commercial, or hospital envi‐
ronment.
IEC
For use in a typical domestic,
60601-1-2:2014,
commercial, or hospital envi‐
Table 9
ronment.
10 V/m
Portable and mobile RF com‐
80 MHz to 6 GHz
munications equipment
39
Electromagnetic Environ‐
ment Guidance