Tool Indications And Warning Signals - Volpi KV100 Mode D'emploi

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 98
Battery should be ke
pt at least at 1.5m from any possible flammable
Maintain the battery; please full charging before storage, charging battery every 6 months when
storing for a long time.
Don't operate switch at storage.

17 Tool indications and warning signals

17.1 Indication of charger
LED
Three squares green
Two squares green
One squares green
One squares red
17.2 Warning tone of the battery
Warning
Interval
note
1 time
0.3 second
2 times
0.3 second
3 times
0.3 second
4 times
0.3 second
4 short
0.1 second
note
10 times
0.3 second
1 long
5 second
note
Li-Battery
Battery capacity
80%-100%
50%-80%
30%-50%
Below 30%
Charger
Indication
Red
Green
Warning
Situation or Problem
Frequency
Indicate: start up and shutdown, small and big
No repeat
blade mouth switch
Indicate: is charging or shear is not connected
well, please check if the nuts at both ends of the
10 repeats
shear and battery is connected firmly.
Indicate: open hall signal of stroke plate faults
10 repeats
Send the shear to Customer Assistance
Indicate: closed hall signal of stroke plate faults
10 repeats
Send the shear to Customer Assistance
The battery is used up.
No repeat
battery soon.
The scissor has worked for 100,000 times
already. Please remove the front cover, blade,
10 repeats
pin, lead screw and re-grease the connection
joints.
Scissor has worked for 500,000 times, and it
10 repeats
should be inspected and maintained fully.
Please contact your local dealer for servicing.
substance.
Voltage
34 V <Vcc ≤ 42 V
Vcc ≤ 34 V
Situation
Charging
Charging finished
Please charge the
35

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Kv150Kv200

Table des Matières