for Intubated Patients
Instructions for use:
1. Firmly connect the small-end (male) of the Microstream
Adapter
1
source).
2. Firmly connect the patient's endotracheal tube connector into
the large-end (female)
3. When using a VitaLine™ with an incubator, string the VitaLine™
tubing through the tubing access door of the incubator such that
one Nafion
while the second Nafion
outside the incubator, close to the incubator tubing access door.
Ensure that the tubing is not twisted or crimped.
4. Twist the female luer connector
Microstream
5. Following connection of the CO
values appear on the monitor display.
6. To remove the airway adapter from the breathing circuit, twist
the airway adapter outwards by a quarter turn while firmly
gripping the closest section of the ventilation source or patient's
endotracheal tubing, and disconnect.
7. For use with a closed suction system, see sketches on the
following pages.
1
3
6
1. Male-end of airway adapter
2. Female end of airway adapter
3. Humidity filter in incubator
4. Humidity filter outside incubator
5. Female luer connector
6. Microstream
®
to the female-end of the Wye piece (ventilation
2
* humidity filter
®
input port
6
®
Legend
input port
®
EtCO
2
of the Microstream
3
is positioned inside the incubator,
* humidity filter
®
securely but gently into the
5
.
sampling line, check that CO
2
4
Sampling
®
Airway Adapter.
®
4
is positioned
2
5
Airway
2
3