Owner's Guide
•
5°C (41°F)
•
50%@40°C (104°F)
• Main supply voltage fluctuations not
to exceed ± 10% of the nominal
voltage.
• Transient over voltages category II.
• Pollution degree 2.
• Class I equipment.
X.
Cautions and Warnings
NOTE: Failure to comply with these
Caution and Warning statements may
result in injury to personnel or equipment
damage.
Do not operate the unit
without sufficient solution in
the tank. Do not use water
alone as the carrier bath.
Water is not a satisfactory
coupling agent unless a
wetting agent is used. For best
results use BioSonic General
Purpose, BioSonic Germicidal
Cleaner or BioSonic Enzymatic
Ultrasonic Cleaning solutions.
Do not place objects directly
on bottom of the tank because
they may block cavitation and
cause buildup of heat in the
transducer causing premature
CAUTION!
failure. Use only BioSonic
UC125 accessory baskets,
cassette racks or beakers to
hold items.
Do
not
Provide adequate ventilation
by maintaining at least 2
inches clearance at the back.
Do not subject unit to shock
or impact.
40°C (104°F)
80%@31°C (88°F)
CAUTION!
WARNING!
WARNING!
restrict
airflow.
Do not immerse the unit in
water.
This product is intended for
use only by trained healthcare
professionals.
Use this product in accordance
with the instructions in the
owner's guide.
Unspecified or improper use
of this product may impair
safety protection.
Before
cleaning
items
of
considerable
value
ultrasonically, consult the
item's
manufacturer
for
cleaning precautions.
The power off selection from
the keypad does not turn off
main power to the unit.
This equipment may cause
radio interference or may dis-
rupt the operation of nearby
equipment. It may be neces-
sary to take mitigation mea-
sures such as re-orienting or
relocating the unit or shield-
ing the location.
Never drain tank by manually
discarding solution as this
may cause the unit to stop
working properly.
Use only the supplied power
cord or equivalent.
Do not disassemble the unit.
There are no user serviceable
parts inside.
Keep unit covered to reduce
airborne contaminants.
Do not place fingers or hands
in the ultrasonic tank while in
operation.
Operation for long periods of
time will cause the cleaning
solution to get very hot. To
avoid scalding, do not place
hands in the solution or spill
any onto you. Be careful
when removing an item
from the tank as it may also
be hot.
XI. Order Information
Cat.No
Description
Unit and accessories:
UC125
100 Unit (100V, 50/60 Hz)
115 Unit (115V, 50/60 Hz)
230CE Unit (230V, 50/60 Hz)
230UK Unit (230V, 50/60 Hz)
240 Unit (240V, 50/60 Hz)
UC155XD Accessory Kit
3/4 FingerGuard Basket, (2) Beakers,
(2) Covers, (2) Positioning Rings and (2)
Beaker Holders
UC151XD Full Size FingerGuard Basket
UC152XD 3/4 FingerGuard Basket
UC153
One Beaker Holder
UC350
Replacement Drain Filters
UC360
Replacement Drain Hose
UC53
600 ml Beaker with Cover and
Positioning Ring
Tank Solutions:
UC30
General Purpose Cleaning Solution
Concentrate in MeterDose
Bottle (16 oz.)
UC32
Enzymatic Ultrasonic Cleaner
Concentrate in MeterDose
Dispensing Bottle (8 oz.)
UC34
Plaster & Stone Remover
Powder Concentrate in
Jar (1.85 lbs.)
UC38
Germicidal Ultrasonic Cleaning
Concentrate in MeterDose
Dispensing Bottle (8 oz.)
UC39
Cement Remover Concentrate
in MeterDose Dispensing
Bottle (16 oz.)
UC42
Germicidal Ultrasonic Cleaning
Concentrate in MeterDose
Dispensing Bottle (16 oz.)
XII. Maintenance
A. Cleaning
All components should be cleaned
each day by wiping with a soft cloth
moistened with a commercially available
mineral deposit remover followed by
cleaning with isopropyl alcohol or a
mild disinfectant.
Do not use petroleum
based solvents, iodophors
or phenolic based products.
(Iodophors and phenolics
can stain the surface of the
unit.) Clean up all liquid
spills immediately. Rust
CAUTION!
deposits on tank can be
cleaned/removed with a
commercially
stainless steel cleaner.