Riester Praxisdienst EliteVue Mode D'emploi page 15

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 18
Directives and manufacturer's declaration - Electromagnetic immunity
The EliteVue instrument is intended for use in the electromagnetic environment specified below. The custo-
mer or user of the EliteVue should ensure that it is used in such an environment.
IEC 60601 test
Immunity testing
Electrostatic
Con: ± 8 kV
discharge (ESD)
Air: ± 2,4,8,15 kV
IEC 61000-4-2
Fast transient elec-
trical disturbances/
5/50 ns, 100 kHz,
bursts
±2 kV
IEC 61000-4-4
±0.5 kV voltage
Phase-to-phase
conductor
±2 kV voltage
Surge voltage
Line-to-earth
IEC 61000-4-5
±0.5 kV voltage
outer conductor
outer conductor
±2 kV voltage outer
conductor ground
< 0% UT 0.5 period
IEC 61000-4-11
at 0.45, 90, 135, 180,
Voltage dips, short-
225, 270 and 315
term interruptions
degrees
and
fluctuations in
0% UT 1 period
the supply voltage
and 70% UT 25/30
pursuant to
periods
IEC61000-4-11
Single phase: at 0
degrees (50/60 Hz)
Magnetic field with
30A/m
efficiency-rated
frequencies IEC
50/60 Hz
61000-4-8
NOTE UT is the AC source. Mains voltage before the application of the test level.
Compliance
level
Con: ± 8 kV
Air: ± 2,4,8,15 kV
Not applicable
Not applicable
Not applicable
30A/m
50/60 Hz
Electromagnetic environment -
Instructions
Floors should be wood, concrete or
ceramic tile.
If the floor is covered with synthetic
material, the relative humidity must be
at least 30%.
The quality of the supply voltage should
be that of a typical business or hospital
environment.
The quality of the
supply voltage should be that of a typical
business or hospital environment.
The quality of the supply voltage should
be that of a typical business or hospital
environment.
Mains frequency magnetic fields should
be at a level characteristic of a typical
location in a typical commercial hospital
environment.
15

Publicité

Table des Matières
loading

Table des Matières