Memory (HEME ISO 2000)
General considerations of Saving/Recall.
•
There is space for 350 saved screens, where each saved screen contains a
complete set of parameters that existed at the time of saving.
•
There is no restriction on the size of an individual locations/zones set up in
memory. The only (global) restriction is that the sum of all screens in all the
zones must not exceed the 350 figure.
•
The parameters saved are:
1. Volts function: Uac, Utrms, Udc, U > LIM, U < LIM, Uac rel. offset, Utrms rel.
offset, Udc rel. offset, MIN, MAX, AVE, Frequency, symbol information and
access control.
2. Amps function: similar to volts, but the relative offsets are set to zero.
3. Ω function: R, Rk, R > LIM, R < LIM, MIN, MAX, AVE, symbol information
and access control.
4. Riso function: Riso, Uiso, Iiso, Riso > LIM, Riso < LIM, R
information and access control.
•
Those parameters cannot be viewed which were not on the display at the time
the screen was saved. Thus if a normal screen was saved, it will be recalled as
seen at the time. If a REC screen was saved, only a REC screen can be
recalled. A PC download must be used to view the „hidden" parameters using
the optional SW package WinLOG for HEME ISO 2000.
, I
, symbol
ab
p
- 27 -