4 MAINTENANCE
The QuikRead go instrument has been designed
to be as user-friendly as possible with no need for
regular maintenance. For any repair requirement,
contact your local supplier.
Instrument calibration
The instrument is factory calibrated. Proper func-
tioning of the instrument is checked by the self-
check procedure during every measurement.
In case of malfunction, an error message is dis-
played.
The calibration data defining the overall assay
curve or cut-off value for each test is encoded
on the cuvette label. This information is trans-
ferred automatically to the instrument during every
measurement.
Cleaning the instrument
Periodically clean the exterior of the instrument
using a lint-free cloth dampened with water.
Pay special attention to cleaning the display. Take
care that no liquid runs onto the edges of the dis-
play, the measurement well or the connectors.
If necessary, a mild detergent may be used. Do
not use organic solvents or corrosive substances.
Spillage of potentially infectious material should
be wiped off immediately with absorbent paper tis-
sue and the contaminated areas swabbed with a
standard disinfectant or 70% ethyl alcohol. Materi-
als used to clean spills, including gloves, should
be disposed of as biohazardous waste.
28
MAINTENANCE
Software update
New software can be uploaded to the instrument
with a USB storage, see page 25. Ask your local
supplier for more information.
Changing the clock battery
The instrument has a battery which gives power
to the internal clock. If the clock battery charge is
empty a warning is displayed. The clock battery
can be replaced with a same type of battery (type
CR 2032 3V).
1.
Turn off the instrument (if on).
2.
Unplug the power supply cable.
3.
Place the instrument onto its side on a table.
4.
Open the accumulator unit cover.
5.
If an accumulator unit is in place disconnect
the accumulator connector from the accumu-
lator unit and remove the accumulator unit.
6.
Take the clock battery out of the battery holder.
7.
Place new clock battery (type CR 2032 3V)
into the battery holder text side down.
8.
If an accumulator unit is used, attach the ac-
cumulator connector to the accumulator unit
and press the accumulator unit into place and
ensure that it is placed properly. Close the ac-
cumulator unit cover.
9.
Turn the instrument back to a standing posi-
tion and plug the power supply cable.
10. Start the instrument by pressing the Power
button.
11. Adjust the date and time (Settings –> Personal
settings –> Measurement flow –> Mainte-
nance –> Date and Time).