GMT
GMT is a universal time against which the instru-
ment's time is set. GMT is not visible to the user
but is an internal clock for the instrument.
To set GMT adjust the date and time with the ar-
row buttons.
Internal clock
Date and time have been adjusted to GMT at
the factory.
Date and time are stored in the memory once
a day at start up.
If the clock battery runs out, the clock stops.
When the battery has been replaced, the
clock continues from the last saved time.
The time is adjusted in the Admin settings and
requires a password. A new adjusted time can-
not be before the last saved time. Giving too
old a time will create an error message.
Local time
Adjust the time according to the local time (Set-
tings –> Measurement flow –> Maintenance:
Date & time). The local time is saved to the
memory as an offset to GMT.
When changing the clock battery, adjust the
time and date. The time cannot be adjusted
to more than 24 hours before the last saved
GMT. Giving too old a time prompts an error
message "Setting of date failed. Date is in the
past."
Security settings
When Security settings is set ON, user needs to
log in to QuikRead go always before starting to
use the instrument. Only emergency measure-
26
USAGE
ments can be measured just with an operator
ID, without a password. All users have a role and
need a user account with operator ID, full name
and password. User roles are
Admin:
only one per instrument, operator ID is ADMIN.
The ID is not modifiable.
Supervisor:
has access to all except for performing factory
reset and deleting security logs.
Normal User:
can set user specific settings.
can see QC and LIS offline results, and option-
ally old patient results.
can add new QC lots locally when doing a
Quality control measurement. QC checking
needs to be ON.
can see error log and transfer the log to a USB
Picture 23
Security settings
storage device.
features not accessible are greyed out.
See table on p. 34 of user roles and rights.
To take Security settings into use, go to Settings
–> Measurement flow –> Maintenance –> Admin
settings –> Security settings
Select Security login ON.
Define access to old results.
Admin password can be changed in Manage
admin account. Default admin password is
QRGOSET.
Set up user and supervisor accounts in User
accounts.
Adjust time after a user will be logged out. This
is different to the hibernate settings in Power
save. If the Full standby option in Power save
is in use, the user will be logged out when
Settings
Access to
Security login:
old results:
Admin or
OFF
supervisor
ON
All users
Define security settings.
Cancel
10:32
ADMIN SETTINGS :: SECURITY
Manage admin account
User accounts
Log out timeout:
5 min.
OK