To turn QC checking on, choose
ter the ADMIN password.
QC checking OFF: Quality control checking is
not in use.
QC checking ON: Quality control checking is
in use.
QC lock: If this is chosen, the instrument will
be locked if a quality control result is outside
the critical limits.
To set a new quality control sample, follow the in-
structions below:
1.
Choose Settings –> Measurement flow –>
QC. Give the administrator password.
2.
Choose New lot.
3.
Select the test from the Test list.
4.
Enter the QC lot ID.
5.
Select the unit from the Units list.
6.
Set the Warning range limits.
Settings
MEASUREMENT FLOW :: QC LOT PROPERTIES
CRP
Test:
1526010
QC lot ID:
mg/l
Units:
Warning
38
range
34
Critical range
43
Target value
Cancel
24
USAGE
/ QC and en-
NOTE: Decimals must be marked with a decimal
point rather than a comma.
7.
Set the Critical range limits.
8.
Set the Target value.
9.
Choose Next.
10. Enter the expiration date.
11. Accept twice by pressing OK.
12. Choose Back.
13. Choose Save.
The quality control lot has now been defined.The
parameters of a lot can be edited or the lot deleted
by choosing the line and selecting Edit or Delete.
When starting a Quality control measurement, the
quality control lot can be chosen from the list by
...
choosing
10:04
Abc
to
Abc
48
Abc
Abc
52
Abc
to
Abc
Next
Picture 22
QC lot properties
Printing
Printing OFF: The instrument does not suggest
printing. It is nevertheless possible to print the
result by choosing Print on the Measure/ Result
screen.
Printing ON: After lifting the cuvette the instru-
ment asks "Print current result?". Accept print-
ing by choosing Yes. Cancel printing by choos-
ing No.
Printing ON + Automatic: The instrument prints
every measurement result automatically.
LIS transfer
LIS transfer OFF: The instrument does not
send the results to the Laboratory Information
System.
LIS transfer ON: After printing, if activated,
the instrument asks: "Send result to LIS?"
Accept
sending
by
choosing
Reject by choosing Reject.
LIS transfer ON + Automatic: The instrument
sends the measurement result automatically to
the LIS.
Saving Measurement flow setting
to Profiles for further use
On the main menu screen choose Profiles. Choose
Save as profile, select an empty profile and name
it or select a profile you'd like to modify, give the
profile a new name if needed and choose OK.
Accept.