Pro-care Optima / English
8. Maintenance
8.1 General
1. Check main power cord and plug if there is abrasions or excessive wear.
2. Check mattress cover for signs of wear or damage. Ensure mattress cover and tubes are stubbed
together correctly.
3. Check airflow from the quick connector. The airflow should alternate between two ventilation
tubes in the middle and the other two at the bottom of quick connrector every half-cycle time if
it's in alternating mode.
4. Check the air hoses if there is kink or break. For replacement, please contact local representative
or place of purchase.

8.2 Fuse Replacement

8.3 Air Filter Replacement

1.Disconnect the power plug when a blown fuse is
suspected.
2. Remove the cover of the fuse holder by means
of a flat screwdriver.
3. Insert a new fuse of the correct rating in, and
replace the cover of the fuse holder back. The fuse
should be rated as T1AL/250V type and VDE
approved.
1. Open the air filter cap located at the back of
pump.
2. Replace with a clean filter and put the cover lid
back. The filter is reusable and can be washed
gently with a mild detergent and water. Dry the
filter before use.
3. Check and replace air filter regularly if
environment is dirty. Suggested to do it every six
month.