Pro-care Optima / English
10. Main LCD screen displays the current therapy. Setting can be changed by turning the knob
and the mode which is chosen will be highlighted, then press the knob once to enter or
confirm the setting.
Mode
Head Reposition Mode
5.1 Mode
Alternating
NOTE- When changing between operating modes, the patient's monitoring and
repositioning program should be reviewed.
5.2 Cycle time
10 mins
5.3 Tuning
Original setting
NOTE- The pressure in the air cells can be manually adjusted for patient comfort by
entering the tuning mode in the right side of the LCD screen .
NOTE- You can also check if the pressure is suitable for the patient by sliding one hand
underneath the air cells at the level of the patient's buttocks. Always leave at least 2.54
cm (1 inch) space between patient and the Continuous Low Pressure cell to prevent
bottoming-out.
Cycle time
Shoulder Lifting Mode
Continuous Low Pressure
15 mins
Decrease mattress pressure
Tuning
Mode Choose
Prone Position Timer
Back to previous page
25 mins
20 mins
Increase mattress pressure
16
Back to previous page
Back to previous page