25. Remote control
25.1. Setup for remote control
Remote control
Interface parameters for the circulator are adjusted at configuration level.
Enter or exit the configuration level by pressing the right arrow
enter
at the same time.
The menu item "Atc" appears on the MULTI-DISPLAY. After pressing the
cursor key
, the second menu item "REMOTE" is indicated.
The interface parameters cannot be adjusted unless the display reads
"r 0".
Adjusting interface parameters:
Example: changing the baud rate.
Using the arrow keys
The actual parameter is displayed (example: "br 24" = 2400 bauds).
With the up and down arrows
desired parameter (example: 4800 bauds). The changed numerals are
blinking.
The new parameter is stored in memory by pressing enter
Adjustable interface parameters
REMOTE
BAUDRATE
PARITY
HANDSHAKE
move the cursor to the desired menu.
0 = keypad control mode
1 = remote control mode via RS232C
12 = 1200 bauds
24 = 2400 bauds
48 = 4800 bauds
96 = 9600 bauds
0 = no parity
1 = odd parity
2 = even parity
0 = Xon/Xoff protocol (software handshake)
1 = protocol RTS/CTS (hardware handshake)
change the number to the
.
and