23. ATC - Absolute Temperature Calibration

Thermostat (T
)
T
Measuring point (T
)
M
Measuring point T
M
The ATC function stays activated until resetting to 00.00 °C.
ATC - Absolute Temperature Calibration
ATC serves to compensate a temperature difference that might occur
between circulator and a defined measuring point in the bath tank
because of physical properties.
The difference temperature is determined (∆T= T
as correcting factor (example ∆T = -0.2 K).
• Press the cursor key
• The MULTI-DISPLAY indicates "Atc0".
• With the edit keys
• Using the cursor keys
set the correcting factor and then press enter
(example -0.20 K)
• Press
and
The temperature on the measuring point rises to a temperature of
37.00 °C and is indicated on the MULTI-DISPLAY.
Note:
The correcting always affects the actual working temperature, even if
this is set via the interface
and enter
at the same time.
select "Atc1" and then press enter
and the edit keys
at the same time.