3. Limitations of use
The instrument is designed for pipetting liquids, observing the following
physical limits:
– +10 °C to +40 °C (50 °F to 104 °F) (of instrument and reagent)
– Vapor pressure up to max. 500 mbar.
Aspirate slowly above 300 mbar, in order to prevent the liquid from
boiling.
– Density up to 9.0 g/cm
4. Operating exclusions
Never use the instrument with liquids whose vapors have a corrosive
effect or attack the materials silicone or EPDM.
The instrument is not designed for Pasteur pipettes.
Warning!
Never use or recharge the instrument in an atmosphere with danger
of explosion. Highly flammable liquids (e. g., ether, acetone and
other liquids with a flash point below 0 °C) must not be pipetted.
5. Storage conditions
Store instrument and accessories in a cool, dry place.
Storage temperature:
- 20 °C to + 50 °C (- 4 °F to + 122 °F)
24
3