5.7. Clear all values

Memory full
When the memory has stored 60 results, the display shows «Full M» after a
measurement. From this point onwards, a new measured value is stored by over-
writing the oldest value.
Clear all values
If you are sure that you want to permanently remove all stored values, hold down the
«Memory»-button (the instrument must have been switched off beforehand) until «CL» appears
and then release the button. To permanently clear the memory, press the «Memory»-button
while «CL» is flashing. Individual values cannot be cleared.

6. Error messages/malfunctions

If an error occurs during a measurement, the measurement is discontinued
and a corresponding error code is displayed (Example: Error No. 2).
Erreur n° Cause(s) possible(s)
Err 1
The systolic pressure was determined but afterwards the cuff pressure fell below
20 mmHg. This situation can occur, for example, if the tube becomes unplugged after
the systolic blood pressure has been measured. Further possible cause: No pulse has
been detected.
Err 2
Unnatural pressure impulses influence the measurement result. Reason: The arm was
moved during the measurement (Artefact).
Err 3
Inflation of the cuff takes too long. The cuff is not correctly seated or the hose
connection is not sealed.
ERR 5
The measured reading indicated an unacceptable difference between systolic and
diastolic pressures. Take another measurement following directions carefully. Contact
your doctor if you continue to get unusual readings