7. Error Messages

No.
Error message
E99
System Error, Contact
Services !
E100 Low Battery !
E102 Too Bright !
E103 Too Dark !
E104 No Staff !
E106 Out Of Level !
E108 Data File Error !
E110 Target Too Close !
E111 Target Too far !
E112 Too Cold !
E113 Too Hot !
E114 Invalid Measurement ! Make another measurement. If further measurement proved to be futile,
Error Messages
Counter measure / causes
Hardware faults or file errors or adjustment errors or setting errors
rendering instrument not working properly.
Change to new or freshly charged batteries.
Darken staff or reduce lighting on staff or shade the objective telescope.
Light up staff uniformly.
Check targeting.
Level the instrument.
Data file error.
Move staff or instrument further apart.
Move staff or instrument closer together.
Stop working, external temperature is outside the instrument operating
temperature.
Stop working, external temperature is outside the instrument operating
temperature.
check staff position and Inverse Staff setting, check the lighting condition
at the staff and stray light, check focusing and targeting, check if sufficient
length of barcode in the field of view.