Specifications in SMOOTH and RECORD SMOOTH modes
SMOOTH mode improves signal stability.
Resolution and stability: 1 count.
Filtering time: 3 seconds.

SPECIFICATIONS OF FREQUENCY MEASUREMENTS

On manual ranges:
On the 40.00 A range, the current must be above 4.5 A and on the 400.0 A range, it must be
above 45 A ± 25 % in the 45 Hz - 1 kHz band.
Below 0.5 Hz or if the detection level is too low, the meter reads 0.0 Hz.
Resolution: 1 count.
Stability: 2 counts.
Response time: 500 ms for frequencies above 20 Hz. Up to 4 s for frequencies below
20 Hz.
RANGE
MEASUREMENT RANGE
MEASUREMENT ERROR
On auto range:
The minimum switching threshold is above 300 mA. Between 1800 Hz and 2000 Hz, the
measurement is progressively subject to an instability of from 2 counts to 13 counts.
RANGE
MEASURE-
MENT RANGE
MEASURE-
0.1 % rdg
MENT ERROR
rdg = reading
1 kHz
0.5 Hz to 1 kHz
0.1% rdg + 1 count 0.2% rdg + 1 count
0.5 to 999.9 Hz
0.5 Hz
5.0 Hz
5.0 Hz
999.9 Hz
+1 count
10 kHz
1 kHz to 10 kHz
1000 Hz
1500 Hz
0.2 % rdg
+2 counts
51
1000 to 9999 Hz
1500 Hz
2100 Hz
0.5 % rdg
0.2 % rdg
+5 counts
2100 Hz
9999 Hz
+1 count