Replacing the fuses

To avoid electric shocks, disconnect the test leads from any
source of voltage before removing the battery cover.
1. Disconnect the test leads from the multimeter.
2. Remove the battery cover (two "B" screws) and the battery.
3. Remove the six "A" screws securing the rear panel.
4. Gently remove the old fuse and install the new fuse.
5. Always use a fuse of the proper size and value (0.8A/1000V fast
blow for the 600mA range [SIBA 70-172-40], 10A/1000V fast blow
for the 20A range [SIBA 50-199-06]).
6. Replace and secure the rear panel, battery and battery cover.

5. TECHNICAL FEATURES

Function
Range
Resolution
DC
600 mV
0,1 mV
Voltage
6V
0,001 V
60 V
0,01 V
600 V
0, 1 V
1 000 V
1 V
AC
Voltage
6 V
0,001 V
60 V
0,01 V
600 V
0,1 V
1 000 V
1 V
Accuracy for 5 % to 100 % of range
DC
600 µA
0,1µA
Current
6 000 µA
1 µA
60 mA
0,01 mA
600 mA
0,1 mA
6 A
0,001 A
10 A
0,01 A
(20 A : 30 s max. with reduced accuracy)
AC
Current
600 µA
0,1µA
6 000 µA
1 µA
60 mA
0,01 mA
600 mA
0,1 mA
0,001 A
6 A
10 A
0,01 A
(20 A : 30 s max. with reduced accuracy)
Accuracy for 5 % to 100 % of range
R: reading; D: digits
o
 Accuracy is valid from 18
28
Accuracy
±(0,09 % R + 2 D)
±(0,15% R + 2 D)
50 to 60 Hz
40Hz-1kHz
±(1,0% R
±(2,0% R
+ 3 D)
+ 3 D)
±(1,2% R
±(2,5% R
+ 3 D)
+ 3 D)
±(1,0 % R + 3 D)
800mA/1000V
±(1,5 % R + 3 D)
40 Hz to 1 kHz
±(1,5 % R + 3 D)
800mA/1000V
±(2,0 % R + 3 D)
o
C to 28
C, RH < 75%.
Chapter II
Protection
1 000 V
1 000 V
Fuse HBC
Fuse HBC
10A/1000V
Fuse HBC
Fuse HBC
10A/1000V