9.2 PROBLEMS IN CONNECTION WITH THE LEVEL REGULATION
9.2a The steam boiler is flooded
Problem generated by the
filling valve
* The filling valve flap is no
longer in place and generates
leaks: fur or any other obstacle...
* The filling valve relay got stuck
* Replace gaskets and clean the
filling valve surface
* Replace the electric valve or the
leval box, as necessary.
9.2b The boiler is empty
RISK TO RUIN THE HEATING ELEMENT
The probe is
directly earthed
Check:
- the probe wire
- the valve cable
- the electronic box functioning
- the valve
Replace as necessary
The steam boiler is flooded
Check the state of the heating element
- 19 -
Problem generated by the
water level probe
The probe is covered with an
insulating agent (scale) or the
probe wire is cut out
Clean probes and check
cables
The filling valve is not ON
or its coil is burnt out