8.3c Dosing device safety system
Display of
Display of
cc
co
The 9V power supply of the dosing
The 9V power supply circuit of the
device is short circuited:
dosing device is opened:
Terminal 30 (+) is earthed
Terminal 30 (+) or earth
are disconnected.
To use the unit, press the Continu/Stop key
Check the cables and the connector of the dosing device
Replace the dosing device, as necessary.