SAFETY INFORMATION
This meter has been designed according to EC 61010 concerning
electronic measuring instruments with a measurement category
(CAT IV 600V) and pollution degree 2.
WARNING
To avoid possible electric shock or personal injury, follow these
guidelines:
Do not use the meter if it is damaged. Before you use the meter,
●
inspect the case. Pay particular attention to the insulation
surrounding the connectors.
Inspect the test leads for damaged insulation or exposed metal.
●
Check the test leads for continuity. Replace damaged test leads
before you use the meter.
Do not use the meter if it operates abnormally. Protection may
●
be impaired. When in doubt, have the meter serviced.
Do not operate the meter where explosive gas, vapour or dust is
●
present.
Do not apply more than the rated voltage, as marked on the
●
meter, between terminals or between any terminal and earth
ground.
Before use, verify the meter's operation by measuring a known
●
voltage.
When measuring current, turn off circuit power before connect-
●
ing the meter in the circuit. Remember to place the meter in
series with the circuit.