The filter is produced with (PTFE) hydrophobic material to prevent fluids entering the pneumatic circuit. It should be changed
immediately if it becomes wet or if there is any sign of contamination or discolouration. If should also be changed if the unit is used with
a patient whose risk of contamination is unknown. Don't use the suction unit without the protection filter. If the suction unit is used in
an emergency or in a patient where the risk of contamination is not know the filter must be changed after each use.
Available under request with different versions with complete jar 2000cc.
WARNING: The medical device is provided without a specific suction probe. If this device must be used with a specific suction probe,
the end user is responsible for making sure it complies with the EN 10079-1 regulation.
WARNING: Suction tubes for insertion in the human body purchased separately from the machine should comply with ISO 10993-1
standards on material biocompatibility.
To clean the plastic housing of the device wear disposable latex gloves and clean with denaturated alcohol or hypoclorite solutions.
Washing and / or cleaning the autoclavable jar as to be carried out as follows:
Wear protection gloves and apron (glasses and face mask if necessary) to avoid contact with contaminating substances;
Disconnect the jar from the device
Disconnect all tubes from the jar and the protection filter
Empty and dispose of the content and of the suction catheter according to the laws in force in your country;
Separate all parts of the cover (overflow valve, o-ring);
After disposing of disposable parts and disassembling the jar wash in running cold water and rinse thouroughly.
Then soak in warm water (temperature shall not exceed 60°C). Wash thouroughly and if necessary use a non-abrasive brush to remove
incrustations. Rinse in running warm water and dry all parts with a soft cloth (non-abrasive).
The jar and the cover can be autoclaved by placing the parts into the autoclave and running one sterilization stem cycle at 121°C (1 bar
relative pressure – 15min) making sure that the jar is positioned upsidedown.
Mechanical resistance of the jar is guaranteed up to 30 cycles of sterilization and cleaning at the indicated conditions (EN ISO 10079-1).
Beyond this limit the physical-mechanical characteristics of the plastic may decrease and replacement of the part is therefore
recommended.
After sterilization and cooling at environment temperature of the parts make sure that these are not damaged.
Assemble the jar as follows:
Place the overflow valve into its seat in the cover (under VACUUM connector)
Insert floating valve keeping the o-ring towards the opening of the cage
Place the o-ring into its seat around the cover
After completing assembling operations always make sue that cover seals perfectly to avoid vacuum leackages or liquid
exit
The aspiration tubes can be sterilized on autoclave using a sterilization cycle at 120°C (1 bar relative pressure – 15min. The conical
connector can be sterilized on autoclave using a sterilization cycle at 121°C (1 bar relative pressure – 15min.
The device is ready for a new employment now.
All manuals and user guides at all-guides.com

ACCESSORIES SUPPLIED

COMPLETE ASPIRATION JAR 1000cc
CONICAL FITTING
TUBES SET 8 mm x 14 mm
ANTIBACTERIAL FILTER
EUROPEAN POWER SUPPLY CORD FOR
CIGARETTE LIGTHER CABLE
CLEANING OF ACCESSORIES
DO NOT WASH, STERILIZE OR PUT IN AUTOCLAVE THE ANTIBACTERIAL FILTER