4. Operation
4.10. mV measurement
Resolution: 1 mV / 0.1 mV
Measuring mode:
By stability.
In continuous.
By time.
For more details, see "pH measurement", page 10-11.
Procedure:
10-04-2010
17:15
pH MEASURE
New calibration
in 1 day
mV measurement with printer
See pH measurement with printer, page 12.
Important note:
Measuring mV the temperature is not shown. The temperature is
displayed only if a temperature probe is connected.
mV calibration
Procedure:
10-04-2010
17:15
mV CALIBRATION
14
sensION+ PH31 HACH_LANGE 0710.indd ANG:14
sensION+ PH31 HACH_LANGE 0710.indd ANG:14
All manuals and user guides at all-guides.com
10-04-2010
17:15
mV MEASURE
...by stability
...
Start meas. mV
Measurement in continuous
MEASURING
MEASURING
by stability
in continuous
* * * * mV
510 mV
Sample 1
00:01
Sample 1
etc.
etc.
Modifying standard calibration
Factory Adjust
Calibration mV
Standard 220 mV
...
To A X Value
Start Calibration
Data Introduciton
Factory Adjust
Factory Adjust
Calibration mV
Standard 220 mV
To A X Value
Compensation: 0 mV
Data Introduciton
Factory Adjust
Continue...
CALIBRATING
Calibration mV
standard
Standard 220 mV
To A X Value
* * * *
mV
25˚C
Data Introduciton
Factory Adjust
00:05
CALIBRATING
Calibration mV
standard
Standard 220 mV
To A X Value
235
mV
25˚C
Data Introduciton
Factory Settings
00:06
etc.
Reprogramming mV measurement conditions
Measuring Conditions
Measuring Conditions
Resolution: 1
Meas.:By stability
Limits: No
Start mV meas.
Measuring Conditions
Measuring Conditions
Resolution: 1
Meas.:By stability
Limits: No
Start mV meas.
00:01
Measuring Conditions
Resolution: 1
Meas.:By stability
Limits: No
Start mV meas.
See page 9
Measuring Conditions
Resolution: 1
Meas.:By stability
Limits: No
Start mV meas.
CALIBRATING mV
etc.
Use standard 220 mV
Continue...
CALIBRATING mV
To a X value
etc.
Use specific standard
Continue...
Introducing Data
Data introduction
Compensation?
Manual introduction of calibration data
+000.0 mV
for a sensor previously calibrated
Factory Settings
Compensation: 0 mV
"Factory adjustment" (compensation: 0 mV),
eliminates the compensation in memory.
Resolution mV:
1
0.1
Measurement:
By Stability
In Continuous
etc.
By time
8/7/10 16:45:06
8/7/10 16:45:06