4. Operation
4.9. pH measurement with a connected printer
4.9.1. Printer configuration
The printer can be configured at the start-up of the instrument or afterwards according to:
10-04-2010
17:15
SYSTEM
4.9.2. Calibration
Once the calibration has finished the calibration data will be automatically printed.
The report can be "reduced", "standard" or "GLP" in accordance with the selected option in "System", see page 16.
Reduced report
10-04-2010
17:15
CALIBRATE pH
Technical buffers
...
Start calibration
CALIBRATING pH
Technical Buffers
Use first buffer
Continue...
etc.
Standard and GLP report
10-04-2010
17:15
CALIBRAR pH
10-04-2007
17:15
Operarios
Antonio Sanz
Operario 2
Operario 3
Operario 4
Tampones técnicos
...
Iniciar calibración
CALIBRANDO pH
Tampones técnicos
Use primer tampón
Continuar...
etc.
12
sensION+ PH31 HACH_LANGE 0710.indd ANG:12
sensION+ PH31 HACH_LANGE 0710.indd ANG:12
All manuals and user guides at all-guides.com
SYSTEM
Sample ID
Data Output
Date/Time
Language
Display contrast
SYSTEM
SYSTEM
Sample ID
Data output:
Data Output
Date/Time
Deactivated
Language
For printer
Display contrast
For computer
More options at page 16.
Only if work with users was se-
lected, see "System" page. 16.
SYSTEM
SYSTEM
Data output:
Type of printer
Deactivated
Thermal
For printer
For computer
Dot matrix
Example of calibration reports:
Reduced report.
CALIBRATION REPORT
Thursday, 14 February 2010
----------------------------------------
Technical
Slope
Sens. A.P.
Buffers
mV/pH
----------
-----
----- ---- ---- ---
4.01, 7.00
-59.2
100.3 -0.1 25.3
Calibration frequency, every 24 h.
----------------------------------------
Standard and GLP report.
LABORATORY
TEST
========================================
CALIBRATION REPORT
Thursday, 14 February 2010
----------------------------------------
pH ELECTRODE
Code:
Serial n.:
pHiso.(pH 7):
CALIBRATED 14-02-2008 08:53:00
Technical
Slope
Sens. A.P.
Buffers
mV/pH
----------
-----
----- ---- ---- ---
4.01, 7.00
-59.2
100.3 -0.1 25.3
Calibration frequency, every 24 h.
----------------------------------------
Operator: Pedro Martínez
----------------------------------------
SN:70802 V1.10
08:53
T.
t.
%
mV
'C
s
6
Stirring 25%
SN:70802 V1.10
08:53
50 14T
7890
pH
7.00
T.
t.
%
mV
'C
s
6
Stirring 25%
8/7/10 16:45:04
8/7/10 16:45:04