The bar graph in the display and the scale on the unit in-
dicate the range of the blood pressure which has been
recorded.
If the values for systolic and diastolic pressure are in two
different WHO ranges (e.g. systolic in the high-normal range
and diastolic pressure in the normal range) the graphic WHO
classification on the unit indicates the higher range (high-
normal in the example described).
6. error messages/troubleshooting
In case of faults, the "Er_" message appears in the display.
Error messages can occur when
• the blood-pressure values are unusually high or low (Er1
or Er2 appears in the display),
• the cuff is too tight or too loose (Er3 or Er4 appears in the
display),
• the inflation pressure is higher than 300 mmHg (Er5 ap-
pears in the display),
• inflation takes longer than 3 minutes (Er6 appears in the
display),
• there is a system or device error (ErA, Er0, Er7 or Er8 ap-
pears in the display),
• the batteries are almost empty
In such cases, repeat the measurement. Ensure you do not
move or speak.
If necessary, re-insert or replace the batteries.
All manuals and user guides at all-guides.com
.
15
7. Cleaning and care
• Clean your blood pressure computer carefully using only a
slightly moistened cloth.
• Do not use any detergents or solvents.
• Never hold the instrument under water as otherwise liquid
can penetrate and damage the instrument.
• Never place any heavy objects on the instrument.
8. Technical details
Model no.
BC 44
Measurement
Oscillating, non-invasive blood
method:
pressure measurement on the wrist
Measurement range
Pressure 0-295 mmHg, heart rate
40-180 beats/minute
Precision of
Systolic ± 3 mmHg/
pressure reading
diastolic ± 3 mmHg
pulse ± 5% of displayed value
Dimensions
L 88 x W 66,2 x H 28,4 mm
Weight
Approx. 98 g
Cuff size
For a wrist circumference of approx.
140 to 195 mm
Admissible operating
+5 °C to +40 °C,
temperature
<80 % relative humidity
Admissible
-20 °C to +55 °C,
storage temperature
<95 % relative humidity
Battery life
2 x 1,5 V Micro (Alkaline Type LR 03)