Error Messages/Troubleshooting; Cleaning And Maintenance - Beurer medical BC 57 Mode D'emploi

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 36
Step 4: BC 57:
data transfer immediately fol-
lowing measurements.
®
• If Bluetooth
data transfer
has been activated, data is
transferred after having con-
firmed the user memory and
pressing the START/STOP
button
.
The "HealthManager" app must be active to allow
data transfer.
If your smartphone has a protective cover, remove this
to ensure that there is no interference during the transfer.
Begin the data transfer in the "HealthManager" app.
The Bluetooth
word mark and logos are registered
®
trademarks owned by Bluetooth
of such marks by Beurer GmbH is under license. Other
trademarks and trade names are those of their respec-
tive owners.
Step 4: BC 57:
Data transfer at a later
point:
• Go
to
memory
mode (chapter 7).
Select the desired
user memory. The
transfer
®
Bluetooth
starts
automati-
cally.
SIG, Inc. and any use
®

7. Error messages/troubleshooting

In case of faults, the EE_ message appears in the display.
Error messages may appear if:
• systolic or diastolic pressure could not be measured (
appears on the display)
• systolic or diastolic pressure was outside the measure-
ment range ( or Lo appears on the display)
• the cuff is fastened too tightly or loosely (
on the display)
• the pump pressure is higher than 300 mmHg (
pears on the display)
• pumping up takes longer than 160 seconds (
pears on the display)
• there is a system or device error (
or
appears on the display)
• the data could not be sent via Bluetooth
In such cases, repeat the measurement. Ensure that you
do not move or speak. If necessary, reinsert or replace
the batteries.

8. Cleaning and maintenance

• Clean your blood pressure monitor carefully using a
slightly damp cloth only.
• Do not use any cleaning agents or solvents.