7. Error messages / troubleshooting
In the event of errors, the error message E_ appears on the
display.
Error messages may appear if:
• It was not possible to record the pulse: E 1;
• You move or speak during the measurement: E2 ;
• The cuff is fastened too tightly or loosely: E3;
• Errors occur during the measurement: E4;
• The pump pressure is higher than 300 mmHg: E5;
• The batteries are almost empty
In such cases, repeat the measurement. Ensure that you do not
move or speak. If necessary, reinsert or replace the batteries.
8. Cleaning and storing the device and cuff
• Clean the device and cuff carefully using a slightly damp
cloth only.
• Do not use any cleaning agents or solvents.
• Under no circumstances hold the device and cuff under
water, as this can cause liquid to enter and damage the
device and cuff.
• If you store the device and cuff, do not place heavy objects
on the device and cuff. Remove the batteries.
9. Technical details
Model no.
BC 32
Type
BC32/1
Measurement
Oscillometric, noninvasive blood
method
pressure measurement on the wrist
E6.
19
Measurement range
Cuff pressure 0 – 300 mmHg,
systolic 50 – 250 mmHg,
diastolic 30 – 200 mmHg,
Pulse 40 –180 beats/minute
Display accuracy
Systolic ± 3 mmHg,
diastolic ± 3 mmHg,
pulse ± 5 % of the value shown
Measurement
Max. permissible standard deviation
inaccuracy
according to clinical testing:
systolic 8 mmHg /diastolic 8 mmHg
Memory
2 x 60 memory spaces
Dimensions
L 72 mm x W 69 mm x H 29.2 mm
Weight
Approx. 101 g (without batteries)
Cuff size
135 to 195 mm
Permissible operating
+5 °C to +40 °C, 1593 % relative air
conditions
humidity (noncondensing)
Permissible storage
25 °C to +70 °C, 1093 % relative air
conditions
humidity, 700 –1060 hPa ambient pres
sure
Power supply
2 x 1.5 V
Battery life
For approx. 250 measurements,
depending on the blood pressure level
and/or pump pressure
Accessories
Instruction for use, 2 x 1.5 V AAA batter
ies, storage box
AAA batteries