9. Faults
All manuals and user guides at all-guides.com
9. Faults
Faults
USB data carrier is full
Error on the USB data carrier
The content of the USB data
carrier cannot be read
Error during transmission of
the measurement
Copy process failed
Measurement could not be
saved
Depot is full
Back pressure too high
Back pressure too low
Input pressure too high
Input pressure too low
Depot pressure too high
The capacity of 500
measurements is reached.
WIKA operating instructions analytic instrument, model GA11
Causes
The limit of the USB data carrier's
storage capacity is reached
Access to the USB data carrier
failed.
Error reading the USB data carrier.
Write error on the USB data carrier
Copy process to the USB data
carrier failed.
Write error on the USB data carrier
The intake capacity of the internal
depot is exhausted.
The gas container into which the
measuring gas has been pumped
is under high pressure.
The measured back pressure is
too low.
The minimum pressure is
atmospheric pressure.
The pressure at the inlet is too
high.
The pressure at the inlet is too low.
The pressure in the internal gas
depot is too high.
The capacity of 500
measurements is reached.
Measures
Connect a new USB data carrier or
free up the capacity.
Plug in/pull out the USB data
carrier or use another USB data
carrier.
Plug in/pull out the USB data
carrier or use another USB data
carrier.
Check the USB data carrier via
PC, and format or repair it, if
necessary.
Check the USB data carrier and
repeat the process.
Plug in/pull out the USB data
carrier or use another USB data
carrier.
Empty storage see chapter
5.6 "Emptying depot and
depressurised gas compartment"
Connect a gas container with lower
pressure.
Make a pressure compensation
with ambient pressure.
Use external pressure reducer
Use inlet pressure control unit
for gas analytic instruments (e.g.
model GA05).
Empty storage see chapter
5.6 "Emptying depot and
depressurised gas compartment"
Manually delete the
measurements or activate the
option "Automatic deletion" see
chapter "Settings".
EN
33