8. Replacing sensors
4. Screw in the new sensor in clockwise direction.
Screw in the sensor in a straight manner because it can easily tilt.
Then make sure that the sensor is fully screwed in, in order to prevent leakages and
falsified measuring results.
EN
5. Connect the plug to the sensor.
The plug can only be inserted in one direction. The groove prevents the plug from
being inserted in a wrong way.
6. Reassemble the front panel and switch on the instrument (see step 1).
⇒
The sensor is recognised automatically.
7. Flood the analytic instrument with pure gas.
Perform 3 measurements with pure gas.
32
All manuals and user guides at all-guides.com
WIKA operating instructions analytic instrument, model GA11