Guidance and manufacturer's declaration – electromagnetic immunity
The IRT 6020/6520 is intended for use in the electromagnetic environment specified below.
The customer or the user of the IRT 6020/6520 should ensure that it is used in such an environment.
Immunity test
Electrostatic discharge
(ESD)
IEC 61000-4-2
Radiated RF
IEC 61000-4-3
Conducted RF
IEC 61000-4-6
Electrical fast transient
IEC 61000-4-4
Surge
IEC 61000-4-5
Power frequency magnetic
field
IEC 61000-4-8
Voltage dips, short
interruptions and voltage
variations on power
supply input lines
IEC 61000-4-11
IEC 60601 test
Compliance
level
±6kV Contact
±8kV Air
3 V/m 80MHz to
2.5GHz
Not Applicable
3Vrms 150kHz to
(no electrical
80MHz
±2kV power line
Not Applicable
±1kV I/O lines
±1kV
differential
Not Applicable
±2kV common
3 A/m
>95% dip 0.5 cycle
60% dip 5 cycles
Not Applicable
70% dip 25 cycles
95% dip 5 sec.
Electromagnetic
level
environment – guidance
Floors should be wood,
concrete, or ceramic tile. If
floors are covered with
Complies
synthetic material, the relative
humidity should be at least
Field strengths outside the
shielded location from fixed RF
Complies
transmitters, as determined by
an electromagnetic site survey,
should be less than 3 V/m.
Interference may occur in the
vicinity of equipment marked
with the following symbol:
Separation distance
cabling)
calculation provided above. If a
known transmitter is present
the specific distance can be
calculated using the equations.
The ME equipment
is solely battery powered.
Power frequency
magnetic fields should be at
levels characteristic of a
Complies
typical location in a typical
commercial or hospital
The ME equipment
is solely battery
30%.
environment.
powered.
13