anemic anoxia and toxic anoxia, some patients with serious anemia
may also report good SpO2 measurement.

4. Technical Specifications

A. Display Format: TFT Display
SpO2 Measuring range: 0 ~100 %
Pulse Measuring range: 30bpm ~250bpm
Perfusion Index Measuring Range: 0% ~ 20%
Pulse Wave Display: columniation display and the waveform
display
B. Power Requirements: 2 ×1.5V AAA alkaline battery (or using the
rechargeable battery instead), adaptable range: 2.6V~3.6V.
C. Power Consumption: Smaller than 30mA
D. Resolution: SpO2: 1%, Pulse: 1bpm, Perfusion Index: 0.1%.
E. Measurement Accuracy: ±2% in stage of 70%-100% SpO2, and
meaningless when stage being smaller than 70%. ± 2 bpm during
the pulse rate range of 30-99 bpm and ± 2% during the pulse rate
range of 100~250 bpm
F. Measurement Performance in weak filling condition: SpO2 and
pulse rate can be shown correctly when pulse-filling ratio is 0.4%.
SpO2 error is ±4%, pulse rate error is ±2 bpm or ±2% (select larger).
G. Resistance to surrounding light: The deviation between the value
measured in the condition of man-made light or indoor natural
light and that of darkroom is less than ±1%.
H. It is equipped with a function switch. The Oximeter can be
powered off in case no finger in the Oximeter within 5 seconds.
I. Optical Sensor
Red light (wavelength is 660nm, 6.65mW)
Infrared (wavelength is 880nm,6.75mW)
J. Adjustable alarm range
SpO2 : 0%~100%
Pulse Rate : 0bpm~254bpm

5. Accessories

Hanging rope, pouch, silicone protector, batteries (optional), user
manual
EN