PULOX PO-200A Mode D'emploi page 75

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

NOTE 1 At 80MHz and 800MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations.
Electromagnetic propagation is affected by absorption and reflection
from structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for
radio (cellular/cordless) telephones and land mobile radios, amateur
radio, AM and FM radio broadcast and TV broadcast cannot be
predicted theoretically with accuracy. To assess the electromagnetic
environment due to fixed RF transmitters, an electromagnetic site
survey should be considered. If the measured field strength in the
location in which The Pulse Oximeter is used exceeds the applicable RF
compliance level above, the Pulse Oximeter should be observed to
80MHz to
3
5 .
d
P
E
1
800MHz
800MHz to
7
d
P
E
1
2.5GHz
Where P is the maximum
output power rating of the
transmitter in watts (W)
according to the
transmitter manufacturer
and d is the recommended
separation distance in
meters (m).
Field strengths from fixed
RF transmitters, as
determined by an
electromagnetic site
a
survey,
should be less
than the compliance level
in each frequency range.
Interference may occur in
the vicinity of equipment
marked with the following
symbol:
b

Publicité

Table des Matières
loading

Table des Matières