PULOX PO-200A Mode D'emploi page 74

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

Electrostatic
discharge
(ESD) IEC
61000-4-2
Power
frequency
(50Hz)
magnetic
field
IEC 61000-
4-8
Guidance and manufacture's declaration-electromagnetic immunity
for EQUIPMENT and SYSTEMS that are not LIFE-SUPPORTING
Guidance and manufacture's declaration-electromagnetic immunity
The Pulse Oximeter is intended for use in the electromagnetic
environment specified below. The customer or the user of Pulse
Oximeter should assure that it is used in such an environment.
Immunity
test
Radiated
RF
ICE
61000-4-
3
±6KV
contact
±8KV air
3A/m
IEC60601
Compliance
test level
level
3V/m
80MHz
3V/m
to
2.5GHz
±6KV
contact
±8KV air
3A/m
Electromagnetic
environment -guidance
Portable and mobile RF
communication equipment
should be used no closer to
any part of the Pulse
Oximeter, including cables,
than the recommended
separation distance
calculated from the
equation applicable to the
frequency of the
transmitter.
recommended separation
distance
Floors should be
wood, concrete or
ceramic tile. If floor
are covered with
synthetic material,
the relative
humidity should be
at least 30%.
Power frequency
magnetic fields
should be at levels
characteristic of a
typical location in a
typical commercial
or hospital
environment

Publicité

Table des Matières
loading

Table des Matières