4. Technical Specifications

4.1 Main Performance

A. SpO2 value display
B. PR value and bar graph display
C. PI value display
D. Pulse waveform display
E. Battery power display
F. Low-battery indication: low-battery indication appears before
working abnormity which is due to low-power
G. Review function
H. Screen brightness can be adjusted
I. Volume can be adjusted
J. Display mode can be adjusted
K. PR sound indication
L. With alarm function, the user could set alarm limit.
M. With real-time data uploading function.
N. With clock function
O. With two kinds of power supply mode(alternating current and
internal electrical power source)
4.2 Main Parameters
A. Measurement of SpO2
Measuring range: 0%~100%
Accuracy: When the SpO2 measuring range is 70%~100%,the
permission of absolute error is ±2%, Below 70% unspecified.
B. Measurement of pulse rate
Measuring range:30bpm~250bpm
Accuracy: ±2 bpm or ±2% (select larger)
C. Measurement of PI
Measuring range: 0%~20%
Accuracy: ±2 bpm or ±2% (select larger)
When the PI measure range is 1%~20%,the permission of
absolute error is ±1%
When the PI measure range is 0%~0.9%,the permission of
absolute error is ±0.2%;
D. Resolution
SpO2 : 1%, Pulse rate: 1bpm, PI: 0.1%
E. Measurement Performance in Weak Filling Condition
SpO2 and pulse rate can be shown correctly when pulse-filling
ratio is 0.4%. SpO2 error is ±4%, pulse rate error is ±2 bpm or ±2%
Figure 1