1. Sécurité
12
POUR ASSURER UN LAVAGE CORRECT, IL EST OBLIGATOIRE
D'AJUSTER L'HYDROFLEX PLUS AU TYPE / FABRICANT DE LA
MICROPLAQUE OU DE LA PLAQUE DE BARRETTES UTILISÉE.
C'EST AUSSI VALABLE POUR TOUT FICHIER DE PLAQUE
PRÉDÉFINI, QUI NE CONTIENT QUE DES VALEURS MOYENNES DE
PARAMÈTRES DE PLAQUE QUI DOIVENT ÊTRE VÉRIFIÉS AVEC LE
TYPE DE PLAQUE CORRESPONDANT ET, SI NÉCESSAIRE,
CORRIGÉS AVANT DE METTRE L'HYDROFLEX PLUS EN SERVICE.
SI CETTE PROCÉDURE DE RÉGLAGE N'EST PAS REALISÉE
CORRECTEMENT, IL PEUT EN RÉSULTER D'IMPORTANTS
VOLUMES RÉSIDUELS PAR PUITS AINSI QU'UN LAVAGE
INSUFFISANT DES PUITS ET PEUT SÉRIEUSEMENT AFFECTER LA
POUR PLUS DE DÉTAILS SUR LA FAÇON D'AJUSTER
L'HYDROFLEX PLUS AU TYPE DE MICROPLAQUE OU DE PLAQUE
DE BARRETTES UTILISÉ, VOIR LE CHAPITRE 6 MENU
PARAMETRES DE CE DOCUMENT.
L'APPAREIL EST CONFORME AUX PRESCRIPTIONS D'EMISSION ET
D'IMMUNITE DECRITES DANS LA NORME IEC 61326-2-6 ;
CEPENDANT L'ENVIRONNEMENT ELECTROMAGNETIQUE DOIT
ETRE EVALUE AVANT LA MISE EN FONCTIONNEMENT DE
IL EN EST DE LA RESPONSABILITE DE L'OPERATEUR DE
S'ASSURER DU MAINTIEN D'UN ENVIRONNEMENT
ELECTROMAGNETIQUE COMPATIBLE A L'APPAREIL, ET DE LUI
PERMETTRE AINSI DE FONCTIONNER COMME SOUHAITE.
NE PAS FAIRE FONCTIONNER L'APPAREIL A PROXIMITE
IMMEDIATE DE SOURCES POUVANT GENERER DE FORTES
RADIATIONS ELECTROMAGNETIQUES (EX : DES SOURCES
D'EMISSION RF INTENTIONNELLES NON BLINDEES) CAR CELLES-
CI PEUVENT NUIRE AU BON FONCTIONNEMENT DE L'APPAREIL ET
ENGENDRER DES RESULTATS FAUSSES.
FLACON DE REJETS – NIVEAU DU LIQUIDE
S'ASSURER QUE LE NIVEAU DE LIQUIDE DU FLACON DE REJETS
EST TOUJOURS MAINTENU SOUS LE NIVEAU MAXIMUM INDIQUÉ
SUR LE FLACON POUR ÉVITER UN POSSIBLE DÉBORDEMENT. LE
CONTENU DU FLACON DE REJETS ÉTANT POTENTIELLEMENT
INFECTIEUX, PORTER DES VÊTEMENTS DE PROTECTION (GANTS,
BLOUSE ET LUNETTES DE PROTECTION) LORS DU
VIDAGE/MANIPULATION D'UN FLACON DE REJETS
(VOIR 8.3.2 NETTOYAGE DU SYSTÈME DE LIQUIDE - NETTOYAGE
DES FLACONS DE REJETS).
SE RENSEIGNER SUR LES LIEUX DE COLLECTE ET LES
MÉTHODES DE REJET APPROPRIÉES DANS VOTRE PAYS, ÉTAT
HYDROFLEX PLUS N° Doc. 30199908 N° Version 1.1
AVERTISSEMENT
QUALITÉ DES ESSAIS.
AVERTISSEMENT
L'APPAREIL.
AVERTISSEMENT
OU RÉGION.
2022-01