GB
Checking the device
AbS
A530
A530
Mode
Zero
Test
wavelength
wavelength
0.0.0
Zero
Test
wavelength
wavelength
RESULT
RESULT
Verifikationsstandard-Kit
The verification standards for the MD 100 / MD 110 / MD 200 serveto check the accuracy
and reproducibility of measurement results based on the integrated wavelengths.
The kit includes a blank and 6 different measuring vials to check 6 different wavelengths.
The verification standard kit thus enables the inspection of all MD 100 / MD 110 / MD 200
seriesphotometers.
The shelf life of the standard is two yearsfrom the date of manufacture, provided it is used
and stored correctly. The measurements are made in units of mAbs.
Part-
Material
Number
21 56 70
Verifikationsstandard-KitMD 100 / MD 110 / MD 200
46
The following appearsbriefly on the display:
AbS followed by the wavelength to be measured (e.g. A530 for
530 nm).
For deviceswith severalLEDs,selectthe desired wavelength by pressing
the [Mode] button severaltimes as needed (scroll).
Position the vial marked "Zero" in the sample chamber. Placethe seal
cap and pressthe [Zero/Test] button.
flashes for approx. 3 seconds
Confirm the zero adjustment
Swirl the vial marked with the selected wavelength severaltimes and
position it in the sample chamber. Closewith the rubber sealcap and
press [Zero/Test].
flashes for approx. 3 seconds
Compare the displayedresult with the value printed on the certificate,
taking into account the tolerances (seealso notes on the certificate).
If the result falls outside of the tolerance and the standardsare within
the expiration date, we recommend returning the device to the ma-
nufacturer or distributor. It is useful to have the manufacturer perform
an inspection, as this means the device is checked for technical faults
in addition to the calibration.
MD1X0/200_1b 09/2020